National standard of the units of kerma in air and of kerma rate in air

The standard is produced and is in operation since 2001

 

Standard composition:

  • dosage meter UNIDOS 1001 with an ionization chamber M30001;
  • dosage meter UNIDOS with ionization chambers M32002 and M32003;
  • ionization chamber for x-ray measurements BABYLINE-88;
  • x-ray apparatus PANTAK 320HF with a set of filters of «NARROW SPECTRUM» series according to standards ISO 4037 and a set of lead collimators;
  • barometers;
  • thermometer ТЛ-4;
  • humidity meter 5804;
  • measuring buggy with accessories;
  • laser systems of positioning;
  • irradiation machine OB-2;
  • irradiation machine OB-6;
  • panoramic irradiator OB-34;
  • phantom MR 5101-T with an adaptor MR 5101A-1;
  • control source of strontium-90 for chamber М30001;
  • control source of strontium-90 for chambers М32002 and М32003;
  • the system of video camera surveillance;
  • light warning system indicating activation of irradiation source;
  • contact blocking system of the sources of irradiation;
  • the system of blocking sources of irradiation with the help of infrared barriers;
  • conditioning systems and the system of temperature control;
  • uninterruptible power supply unit 1 kVA

Metrological characteristics:

  • range of kerma in the air – 7•10–9 - 3•106 Gray;
  • range of kerma rate in the air – 7•10–10 - 45 Gray/s;
  • range of photonic energy irradiation – 33 - 1250 keV;
  • value of expanded uncertainty of kerma rate measurement in the air – 2,2 - 7,0 %;
  • долговременная нестабильность за год составляет не более 0,3 %

BelGIM calibration and measurement capabilities (СМС) in the field of ionization radiation are included into Attachment С of the Agreement on mutual recognition of national standards, calibration and measurement certificates (CIPM MRA) and published on International Bureau of Weights and Measures website

СМС ionizing radiation (kerma)

Sources of traceability:

Russia, FSUE “All-Russian Research Institute of Metrology named after D.I. Mendeleyev”, State primary standard of the units of exposure dose, exposure rate and the stream of x-ray and gamma-radiation (GET 8-82)

Scope of application:

The standard allows to carry out research, verification, calibration of devices, measuring:

  • kerma in the air; kerma rate in the air; exposure dose;
  • exposure dose rate; dose, absorbed in water;
  • dose rate, absorbed in water; equivalent dose;
  • equivalent dose rate;
  • other dose characteristics of background radiation

 