У ВАС отключен Java Script. Полноценное функционирование сайта НЕ ОБЕСПЕЧЕНО

+375 17 288-09-38 (Fax)
info@belgim.by

Reference standard of the mean laser radiation power unit

The standard is created and is in operation since 2006

Reference standard of the mean laser powerReference standard of the mean laser power

Standard composition:

  • a set of devices for measuring the mean laser power;
  • a set of devices for transfer of the size of mean laser power unit;
  • a set of devices for controlling, collecting and processing measurement information

Metrological characteristics:

  • fixed lengths of waves of laser radiation reproducing — 0,532 and 10,6 mcm;
  • the range of mean power of continuous laser radiation — from 0,05 to 1,3 W;
  • SE of reproduction, not more than 0,4 %;
  • transmission error of the unit size by the verified (calibrated) means of measurement, presented in the form of SE, not more than 0,5 %

Sources of traceability:

Russia, FSUE «All-Russian Research Institute of optical and physical measurements», State primary standard of the mean laser radiation power unit (GET 28-91)

 Scope of application:

  • testing, verification and calibration of means of calibration of mean laser radiation power;
  • public health (for measuring power of radiation of laser therapeutic and other equipment);
  • industry;
  • labour protection in terms of laser radiation

 